GDS-200 & GDS-300 Programming Manual

88

Parameter

Reference waveform 1, 2

Horizontal co-ordinates

Return parameter Returns the reference waveform position

Example

:REF1:TIMebase:POSition -5.000E-5
Selects reference 1, and then sets the horizontal
position to -50us.

:REF:TIMebase:SCALe

Description

Sets or returns the selected reference waveform
time base scale.

Syntax

:REF:TIMebase:SCALe { | ?}

Related
commands

:REF:DISPlay

Parameter

Reference waveform 1, 2

Horizontal scale

Return parameter Returns the reference waveform scale.

Example

:REF1:TIMebase:SCALe 5.00E-4
Selects reference 1, and then sets the horizontal
scale to 500us/div.

:REF:OFFSet

Description

Sets or returns the selected reference waveform
vertical position (offset).

Syntax

:REF:OFFSet { | ?}

Related
commands

:REF:DISPlay

Parameter

Reference waveform 1, 2

Vertical offset

Return parameter

Returns the reference waveform vertical
position.

Set

Query

Set

Query