beautypg.com

Dmm:cursorpos – GW Instek GDS-200 Series Programming Manual User Manual

Page 130

background image

GDS-200 & GDS-300 Programming Manual

130

parameter

TIME3M

TIME6M

TIME15M

TIME30M

TIME1H

TIME3H

TIME6H

Example

:DMM:TIMe 6M
Sets the measurement interval time to 6M.

:DMM:CURSORPOS

Description

Sets the cursor position for trend plot in the
Advanced DMM function.
This command is only applicable when the Hold
function is turned on.

Syntax

:DMM:CURSORPOS { | ? }

Parameter

0 ~ 400

Return parameter

Returns the cursor position (#data);
time at position; measurement value
at position

Example

DMM:HOLD ON
DMM:CURSORPOS 200
:DMM:CURSORPOS?
200;20.0s;0.0000V
Returns cursor position #200 @ 20s with a value of
0V.

Set

Query