Measure:pduty, Measure:ppulse – GW Instek GDS-200 Series Programming Manual User Manual
Page 72
GDS-200 & GDS-300 Programming Manual
72
Chan Off
Indicates the source channel is not
activated.
Note
Before using this command, select the
measurement channel. See the example below.
Example
:MEASure:SOURce1 CH1
:MEASure:NWIDth?
4.995E-04
Selects Channel 1, and then measures the negative
pulse width.
:MEASure:PDUTy
Description
Returns the positive duty cycle ratio as percentage.
Syntax
:MEASure:PDUTy?
Related
commands
:MEASure:SOURce
Return parameter
Returns the positive duty ratio.
Chan Off
Indicates the source channel is not
activated.
Note
Before using this command, select the
measurement channel. See the example below.
Example
:MEASure:SOURce1 CH1
:MEASure:PDUTy?
5.000E+01
Selects Channel 1, and then measures the positive
duty cycle.
:MEASure:PPULSE
Description
Returns the number of positive pulses.
Syntax
:MEASure:PPULSE?
Query
Query