beautypg.com

Measure:pduty, Measure:ppulse – GW Instek GDS-200 Series Programming Manual User Manual

Page 72

background image

GDS-200 & GDS-300 Programming Manual

72

Chan Off

Indicates the source channel is not
activated.

Note

Before using this command, select the
measurement channel. See the example below.

Example

:MEASure:SOURce1 CH1
:MEASure:NWIDth?
4.995E-04
Selects Channel 1, and then measures the negative
pulse width.

:MEASure:PDUTy

Description

Returns the positive duty cycle ratio as percentage.

Syntax

:MEASure:PDUTy?

Related
commands

:MEASure:SOURce

Return parameter

Returns the positive duty ratio.

Chan Off

Indicates the source channel is not
activated.

Note

Before using this command, select the
measurement channel. See the example below.

Example

:MEASure:SOURce1 CH1
:MEASure:PDUTy?
5.000E+01
Selects Channel 1, and then measures the positive
duty cycle.

:MEASure:PPULSE

Description

Returns the number of positive pulses.

Syntax

:MEASure:PPULSE?

Query

Query