beautypg.com

Measure:rise, Measure:fall – GW Instek GDS-200 Series Programming Manual User Manual

Page 70

background image

GDS-200 & GDS-300 Programming Manual

70

Note

Before using this command, select the
measurement channel. See the example below.

Example

:MEASure:SOURce1 CH1
:MEASure:PERiod?
1.0E-3
Selects Channel 1, and then measures the period.

:MEASure:RISe

Description

Returns the first pulse rise time.

Syntax

:MEASure:RISe?

Related
Commands

:MEASure:SOURce

Return parameter

Returns the rise time.

Chan Off

Indicates the source channel is not
activated.

Note

Before using this command, select the
measurement channel. See the example below.

Example

:MEASure:SOURce1 CH1
:MEASure:RISe?
8.5E-6
Selects Channel 1, and then measures the rise time.

:MEASure:FALL

Description

Returns the fall time measurement result.

Syntax

:MEASure:FALL?

Related
Commands

:MEASure:SOURce

Return parameter

Chan Off

Indicates the source channel is not
activated.

Query

Query