Measurement:meas:maximum – GW Instek GDS-200 Series Programming Manual User Manual
Page 83

COMMAND DETAILS
83
Return parameter MEAS
from 1 to 4.
Note
The measurement source(s), measurement number
and measurement type must first be set before a
measurement result can be returned.
Example
:MEASUrement:MEAS1:SOUrce1 CH1
:MEASUrement:MEAS1:TYPe PK2PK
:MEASUrement:MEAS1:VALue?
5.000E+0
Selects channel 1 as the source for measurement 1
and sets measurement 1 to peak to peak
measurement. The result returns the peak to peak
measurement.
:MEASUrement:MEAS
Description
Returns the maximum measurement results for the
selected measurement from the last time the
statistics were reset. This is a statistics related
command.
Syntax
:MEASUrement:MEAS
Related
Commands
:MEASUrement:STATIstics:MODe
Parameter
MEAS
from 1 to 4.
Example
:MEASUrement:MEAS3:SOUrce1 CH1
:MEASUrement:MEAS3:TYPe PK2PK
:MEASUrement:STATIstics:MODe ON
:MEASUrement:MEAS3:MAXimum?
2.800E-02
Returns the maximum measurement result for
measurement number 3.
Query