beautypg.com

Measure:high, Measure:low – GW Instek GDS-200 Series Programming Manual User Manual

Page 62

background image

GDS-200 & GDS-300 Programming Manual

62

Chan Off

Indicates the source channel is not
activated.

Note

Before using this command, select the
measurement channel. See the example below.

Example

:MEASure:SOURce1 CH1
:MEASure:AMPlitude?
3.76E-3
Selects Channel 1, and then measures the
amplitude.

:MEASure:HIGH

Description

Returns the high voltage/current.

Syntax

:MEASure:HIGH?

Related
Commands

:MEASure:SOURce

Return parameter

Returns the high value.

Chan Off

Indicates the source channel is not
activated.

Note

Before using this command, select the
measurement channel. See the example below.

Example

:MEASure:SOURce1 CH1
:MEASure:HIGH?
3.68E-3
Selects Channel 1, and then measures the high
voltage/current.

:MEASure:LOW

Description

Returns the low voltage/current.

Syntax

:MEASure:LOW?

Query

Query

This manual is related to the following products: