beautypg.com

Measurement:statistics:mode, Measurement:statistics – GW Instek GDS-200 Series Programming Manual User Manual

Page 85

background image

COMMAND DETAILS

85

Example

:MEASUrement:MEAS3:SOUrce1 CH1
:MEASUrement:MEAS3:TYPe PK2PK
:MEASUrement:STATIstics:MODe ON
:MEASUrement:MEAS3:MINImum?
1.600E-02
Returns the minimum measurement result for
measurement number 3.

:MEASUrement:STATIstics:MODe

Description

Puts the statics measurement results on the display
or queries whether the statistics are displayed. This
is the equivalent of showing the automatic
measurements in the expanded mode (statistics
mode on) condensed mode (statistics mode off).

Syntax

:MEASUrement:STATIstics:MODe {OFF | ON | ?}

Related
commands

:MEASUrement:STATIstics

Parameter/
Return parameter

ON

Display the statistics on the screen.

OFF

Remove the statistics from the screen

Example

:MEASUrement:STATIstics:MODe ON

Displays statistics on the screen.

:MEASUrement:STATIstics

Description

Resets the statics calculations. This command will
clear all the currently accumulated measurements.

Syntax

:MEASUrement:STATIstics {RESET}

Set

Query

Set

This manual is related to the following products: