beautypg.com

Watlow Series F4S/D User Manual

Page 41

background image

Wa t l o w S e r i e s F 4 S / D

P r o f i l e P r o g r a m m i n g

4 . 9

88

90

92

94

98

100

102

104

106

96

3

Steps

6

7

8

9

10 11

13

12

14 15 16 17 18 19

17 18 19 20 21 22 23

6

7

8

9

10 11 12 13 14 15

0

1

2

3

4

5

16

Test Period in Hours

58

60

62

64

68

70

74

76

78

80

82

86

66

72

84

88

1 2

% Relative Humidity

Relative Humidity

Temperature ( F)

Temperature ( F)

20

21

End of test

Step 20 jumps to Step 3 and repeats 20 times.

Step

Channel 1

Channel 2

4

5

Figure 9a — Profile Chart for Military Standard 810D Test.

Figure 9b — Graph of Military Standard 810D Test.

Step Step

Date/Day, Wait

Set Events

Time

Rate

Set

Set

PID

Guar.

Jump to

Step

Repeats End

Nmbr Type

Time

for

1 2 3 4 5 6 7 8

H M S

Pt 1

Pt 2

Set

Soak

Profile

Step

1

Ramp Time

1 sec.

88°F

88%

2

Soak

Process1&2

1 sec.

3

Soak

5 hrs.

4

Ramp Time

1 hr.

90°F

85%

5

Ramp Time

1 hr.

93°F

80%

6

Ramp Time

1 hr.

96°F

76%

7

Ramp Time

1 hr.

98°F

73%

8

Ramp Time

1 hr.

100°F 69%

9

Ramp Time

1 hr.

102°F 65%

10

Ramp Time

1 hr.

104°F 62%

11

Ramp Time

1 hr.

105°F 59%

12

Soak

Process1&2

3 hrs.

13

Ramp Time

1 hr.

102°F 65%

14

Ramp Time

1 hr.

99°F

69%

15

Ramp Time

1 hr.

97°F

73%

16

Ramp Time

1 hr.

94°F

79%

17

Ramp Time

1 hr.

91°F

85%

18

Ramp Time

1 hr.

90°F

85%

19

Ramp Time

1 hr.

89°F

88%

20

Jump

3

20

21

End

All Off

2

Dual

Channel

F4D