Calibration menu map, Retransmit 2 setup and calibration – Watlow Series F4S/D User Manual

Page 112

background image

9 . 6

C a l i b r a t i o n

Wa t l o w S e r i e s F 4 S / D

<

to adjust the display to the reading on the

volt/ammeter. The controller should stabilize

within one second. Repeat until the volt/amme-

ter reads 4.000mA. Press . to store the value.

4. Press the Right Key . at the Calibrate

Rexmit 1 prompt. At the 20.000mA prompt

press . once. Use the Up Key > or the Down

Key < to adjust the display to the reading on

the volt/ammeter. The controller should stabi-

lize within one second. Repeat until the

volt/ammeter reads 20.000mA. Press . to

store the value.

Volts
5. Connect the volt/ammeter to terminals 48 (+)

and 49 (-).

6. Press the Right Key . at the Calibrate

Rexmit 1 prompt. At the 1.000V prompt press
.

once. Use the Up Key > or the Down Key

<

to adjust the display to the reading on the

volt/ammeter. The controller should stabilize

within one second. Repeat until the volt/amme-

ter reads 1.000V. Press . to store the value.

7. Press the Right Key . at the Calibrate

Rexmit 1 prompt. At the 10.000V prompt press
.

once. Use the Up Key > or the Down Key

<

to adjust the display to the reading on the

volt/ammeter. The controller should stabilize

within one second. Repeat until the volt/amme-

ter reads 10.000V. Press . to store the value.

8. Rewire for operation and verify calibration.

Retransmit 2 Setup and Calibration

1. Connect the correct power supply to terminals

1, 2 and 3 (see the Wiring Chapter and the

Appendix).

Milliamperes
2. Connect the volt/ammeter to terminals 47 (+)

and 46 (-).

3. Press the Right Key . at the Calibrate

Rexmit 2 prompt. At the 4.000mA prompt press
.

once. Use the Up Key > or the Down Key

<

to adjust the display to the reading on the

volt/ammeter. The controller should stabilize

within one second. Repeat until the volt/amme-

ter reads 4.000mA. Press . to store the value.

4. Press the Right Key . at the Calibrate

Rexmit 2 prompt. At the 20.000mA prompt

press . once. Use the Up Key > or the Down

Key < to adjust the display to the reading on

the volt/ammeter. The controller should stabi-

lize within one second. Repeat until the

volt/ammeter reads 20.000mA. Press . to

store the value.

Volts
5. Connect the volt/ammeter to terminals 45 (+)

and 46 (-).

6. Press the Right Key . at the Calibrate

Rexmit 2 prompt. At the 1.000V prompt press
.

once. Use the Up Key > or the Down Key

<

to adjust the display to the reading on the

volt/ammeter. The controller should stabilize

within one second. Repeat until the volt/amme-

ter reads 1.000V. Press . to store the value.

7. Press the Right Key . at the Calibrate

Rexmit 2 prompt. At the 10.000V prompt press
.

once. Use the Up Key > or the Down Key

Calibration Menu Map

Calibrate Input 1
Calibrate Input 2
Calibrate Input 3
Calibrate Output 1A
Calibrate Output 1B
Calibrate Output 2A
Calibrate Output 2B
Calibrate Rexmit 1
Calibrate Rexmit 2
Restore In1 Cal
Restore In2 Cal
Restore In3 Cal