13 transfer of data from system ii to xds, Transfer of data from system ii to xds, Transfer of calibration data on the same computer – Metrohm Vision Manual User Manual

Page 86

background image

84

▪▪▪▪▪▪▪

Click on Browse button and in the new window find the project location.

Select the ProjData.D01 file and click on Save. Vision defaults the Project ID to the name of

folder with project files. Modify it if necessary. In the Destination Directory field enter the
location of copied project. Enter optional description and click on OK.

3.13

Transfer of Data from System II to XDS

3.13.1

Transfer of Calibration Data on the same computer

A calibration equation developed on a System II instrument cannot be used to predict on an

XDS instrument. The spectra from the System II can be used (after conversion) to create a new
calibration for XDS predictions.

Before leaving the System II instrument, while in the project you want to convert for use on

the XDS instrument:

Go to Configure/Options and select the option for Reference Standardization followed by

Acquire/Connect

At the Select Data Collection Method dialogue box, create a new DCM for your instrument.

The box for Reference Standardize should be active. Select that DCM followed by OK.

Move to Diagnostics/Reference Standard/Create Reference Standard

Vision prompts for the ceramic reference. Position the ceramic reference and click on OK for

scanning to begin.

Vision requests positioning of the 80% Reflectance Reference. Position the correct reference

standard for your instrument and click OK to scan the Certified Reference.

Vision then requests the Standard File for the Certified Reference. This file is on the diskette

accompanying the Standard(s) and may be copied to the Vision directory for ease of use.
Insert the diskette into drive A and access the file. The file is contained in RSSxxxx.da Click on
the file, then click Open.