8 editing vision configuration, Editing vision configuration – Metrohm Vision Manual User Manual

Page 56

background image

54

▪▪▪▪▪▪▪

Standardization.

Blank Correction. This option is used to create a correct transmission reference on the

XDS Rapid Liquid Analyzer

Master Standardization. This method is typically not used. Do not select.

Use Instrument Calibration (XDS only). This is a method used to adjust the instrument

wavelength profile to an external, traceable wavelength standard. This feature must
be used on the XDS instruments to assure method transferability.

Use Window Correction (XDS only). This is selected when using the XDS Process

instrument with either Transmission Pair or Interactance Immersion Probes. This
option should not be used with Interactance Reflectance.

Click on OK when done.

Some options are available on XDS instruments only.

3.8

Editing Vision Configuration

Edit Configuration window allows to select visible elements of Vision interface, lamp management,
lamp status for XDS instruments, and logout interval.

Select Configure/Configuration from main menu. For interface options, click on View tab:

Select desired interface elements: Status Bar, Vision Toolbar, Tooltips, and Time. All elements

are active after Vision installation by default.

The default Gray color scheme gives Vision appearance consistent with most Windows

applications. It can be changed to White to improve appearance of spectra pasted to
presentations and documents.

For logout configuration and lamp management click on Logout Interval/Lamp Life tab: