Master Standardization versus Reference Standardization

Vision provides two standardization methods: Reference Standardization and Master Standardization.
Reference Standardization is described in the Diagnostics section of this manual.

To avoid confusion the following table lists major differences between Reference Standardization and
Master Standardization:

Property

Reference Standardization Master

Standardization

Supported by Metrohm
NIRSystems instruments

Laboratory reflectance
systems, XDS Process

All systems

Standardization algorithm

Photometric offset correction
using a suitable standard to
achieve "absolute
reflectance".

Photometric offset correction
at any reflectance or
transmittance level with
optional scaling

Standardization correction
storage

In the analyzer's PROM

In the Project or Library
database

Standardization type

Global, one correction for all
methods marked as reference
standardized

Local, various products or
projects can have different
corrections

NIST traceable

Yes

No

Allows to set up your own
master instrument

Not Necessary

Yes

Data Collection Method

DCM’S are marked as such:
standardized or not-
standardized.

Not dependent on DCM