when the maximum value is within threshold limits. The default threshold is 4.

Correlation in wavelength space

Wavelength correlation is an example of another wavelength method. During the analysis, the dot
product is calculated between the unknown spectrum and the mean spectrum of each library
product. The unknown is identified as a product when the value is higher than the threshold. The
default value of the threshold is 0.85.

4.5.4

Identification Method Development Procedure

Preparation

1.

From the main menu, select Mode/Qualitative Analysis/Identify Method Development to enter

the identification development program.

2.

Open the combo box and double click the library name for which you want to create the

method.

Define the identification parameters

1.

From the menu select Identify/Apply Math.