3 library validation report viewer, Library validation report viewer – Metrohm Vision Manual User Manual
Page 120

118
▪▪▪▪▪▪▪
2.
Click on Start to begin validation. Progress bars show the status of the validation. After the
operation is completed, you can save the results by clicking on “Save” button.
3.
Click on Close to return to the main validation screen.
Viewing the Results
1.
The results screen contains two graph displays and two result displays.
2.
The spectral display graph displays a spectrum after it is clicked on the library tree. The 3-D
bar graph shows the percent of misidentified spectra in each product.
3.
The two result displays show the validation results at the product and sample level.
Printing and saving the validation results
1.
To print the validation report, choose Print Report from the library menu.
2.
Click on Library/Save Results to save the validation report. Type in optional comments and
click on OK button.
4.7.3
Library Validation Report Viewer
The viewer opens the validation report in a separate window, and allows users to save the report into
Excel format or print. To open the viewer click on Library/View Report after the validation is
completed.