3 measurement setting, Settings of demand measurement, Measurement cycle – KYORITSU 6315 Instruction Manual User Manual
Page 61

5.3 Measurement setting
KEW6315
- 59 -
KEW6315
5.3 Measurement setting
Press the
Key.
Change the tabs to “Measurement”.
Settings of demand measurement
“Measurement cycle”
Disable the demand measurement or set the demand measurement cycle in the preset recording period.
When a demand measurement start, the measured demand values will be recorded at the selected
measurement cycle. The cycle time should be selected from the following.
Selection
Not be used./ 10 min/ 15 min/ 30 min
* Default setting is highlighted in gray.
The selected demand measurement cycle has an influence on the selection of the measurement intervals.
Since the measurement interval cannot be set to a longer time than the demand interval, the preset measure-
ment interval may be changed automatically according to the selected demand measurement cycle.
Selectable measurement intervals: 1sec/ 2sec/ 5sec/ 10sec/ 15sec/ 20sec/ 30sec/ 1 min/ 2 min/ 5 min/ 10 min/
15 min/ 30 min.
Move the blue highlight to “Measurement”.
Show the pull-down menu.
Select a desired demand cycle.
Confirm.
Cancel.