KYORITSU 6315 Instruction Manual User Manual
Page 161

11.2 Error messages and actions
KEW6315
- 159 -
KEW6315
the instrument or the sensor.
Message
Detail & Action
Battery level is low.
Powering off
● Connect the instrument to an AC power source, or replace the
batteries with new ones. * Size AA Alkaline battery (LR6) or
fully-charged Size AA Ni-MH battery x 6pcs
See “How to install batteries” (P. 31).
Not having free space on the internal
memory. Format the memory or
delete unnecessary files.
● Check the free space on the internal memory and the number of
the saved files. Max number of the file that can be saved on the
memory is: 3 for measurement data and 8 for the other data. If
the free space is not enough, delete unnecessary files, format
the memory.
See “To delete, transfer or format the recorded data”
(P.
82).
Cannot read the setting file. The file
may be damaged.
● Try again. If still the setting files are not read;
* problems with SD card or KEW6315 are suspected, if the setting
files are on the SD card,
* problems with KEW6315 are suspected, if the setting files
are in the internal memory.
If the problem with KEW6315 is suspected, stop using the
instrument.
Available memory is low.
Check the amount of free space in
the SD card and internal memory.
● Check the free space and the number of saved files on the SD
card and the internal memory. Max number of the file that can be
saved on the memory is: 3 for measurement data and 8 for the
other data. If the space is not enough, delete unnecessary files,
format the card or memory. When using another SD card, it
should be formatted on KEW6315, not on the PC.
See “To delete,
transfer or format the recorded data” (
P.
82).
There is no available space in the
storage area.
Start time is set in the past.
Check the recording start method.
● REC Start is either “Constant rec. / Time period rec.” and the time
set for REC End is set to the past. Check and modify the time and
date. See “(8)/ (9) Setting for recording method” (P. 45).
Failed to start recording.
● Check the “Recording setting” at SET UP menu.
See “
5.4 Recording setting” (P. 71).
● Try again. If still a record does not start, there may be a problem
with either the SD card or the internal memory. Check which is set
as the destination to save the data. If the destination is internal
memory, a problem with KEW6315 is suspected. Stop using the
instrument in this case.
Cannot change the instrument
settings during recording or in
stand-by mode.
● Change of setting is not allowed during a record. To change the
settings, stop record and confirm “Recording stopped.” message
appears and then disappears.