KYORITSU 6315 Instruction Manual User Manual
Page 155

10.4 Specification of Clamp sensor
KEW6315
- 153 -
KEW6315
Rated current
AC 500Arms (707Apeak)
AC 1000Arms (1414Apeak)
Output voltage
AC0 - 500mV
(AC500mV/500A):AC 1mV/A
AC0 - 500mV
(AC500mV/1000A):0.5mV/A
Measuring range
AC0 - 500Arms
AC0 - 1000Arms
Accuracy
(sine wave input)
±0.5%rdg±0.1mV (50/60Hz)
±1.0%rdg±0.2mV (40Hz - 1kHz)
±0.5%rdg±0.2mV (50/60Hz)
±1.5%rdg±0.4mV (40Hz - 1kHz)
Phase
characteristics
within ±1.0°
(5 - 500A/45 - 65Hz)
within ±1.0°
(10 - 1000A/45 - 65Hz)
Temp. & humidity
range (guaranteed
accuracy)
23±5ºC, relative humidity 85% or less (no condensation)
Operating temp.
range
0 - 50ºC, relative humidity 85% or less (no condensation)
Storage temp.
range
-20~60ºC, relative humidity 85% or less (no condensation)
Allowable input
AC500Arms (50/60Hz)
AC1000Arms (50/60Hz)
Output impedance
Approx. 2Ω
Approx. 1Ω
Location for use
In-door use, altitude 2000m or less
Applicable
standards
IEC 61010-1,IEC 61010-2-032
Meas. CAT. III (600V), Pollution degree 2
IEC61326
Withstand voltage
AC5350V/5 sec
Between Jaws – Enclosure, Enclosure – Output terminal, and Jaws – Output terminal
Insulation
resistance
50MΩ or more/ 1000V
Between Jaws – Enclosure, Enclosure – Output terminal, and Jaws – Output terminal
Max conductor size
Approx. ø40mm (max.)
Approx. ø68mm (max.)
Dimension
128(L)×81(W)×36(D)mm
186(L)×129(W)×53(D)mm
Cable length
Approx. 3m
Output terminal
MINI DIN 6PIN
Weight
Approx. 260g
Approx. 510g
Accessory
Instruction manual, Cable marker
Optional parts
7146 (Banana ø 4 Adjuster plug), 7185(Extension cable)