Items to be calculated apparent power s [va, Reactive power q [var, Ps sign – KYORITSU 6315 Instruction Manual User Manual
Page 139

Items to be calculated
KEW6315
- 137 -
KEW6315
Items to be calculated
Apparent power S [VA]
Range
Same as active power.
Displayed digit
Same as active power.
Accuracy
±1dgt against each calculated value (for sum : ±3dgt)
Sign
No polarity indication
Equation
S
c
=
V
c
×A
c;
when P
c
>S
c,
regarding P
c
=
S
c
. c: Measurement channel
1P2W-1 to 4
S
1,
S
2,
S
3,
S
4,
S
sum
=
S
1
+S
2
+S
3
+S
4
1P3W-1 to 2
S
1,
S
2,
S
sum1
=
S
1
+S
2
S
3,
S
4,
S
sum2
=
S
3
+S
4
S
sum
=
S
sum1
+S
sum2
3P3W-2
S
1,
S
2,
S
sum1
=√
3/2( S
1
+S
2
)
S
3,
S
4,
S
sum2
=√
3/2( S
3
+S
4
)
S
sum
=
S
sum1
+S
sum2
3P3W3A
S
1,
S
2,
S
3,
S
sum
=
S
1
+S
2
+S
3 * Phase angles are used.
3P4W
S
1,
S
2,
S
3,
S
sum
=
S
1
+S
2
+S
3
Reactive power Q [Var]
Range
Same as active power.
Displayed digit
Same as active power.
Accuracy
±1dgt against each calculated value (for sum : ±3dgt)
Sign
-
: leading phase (current phase against voltage)
+ (no sign)
: lagging phase (current phase against voltage)
Harmonics reactive power is calculated per ch, and the polarity sign of the
reversed basic waveform is displayed.
Equation
Q
c
=
2
2
c
c
P
S
sign
−
sign: Polarity sign , c: Measurement channel
1P2W-1 to 4
Q
1,
Q
2,
Q
3,
Q
4,
Q
sum
=
Q
1
+Q
2
+Q
3
+Q
4
1P3W(3P3W)-1
to 2
Q
1,
Q
2,
Q
sum1
=
Q
1
+Q
2
Q
3,
Q
4,
Q
sum2
=
Q
3
+Q
4
Q
sum
=
Q
sum1
+Q
sum2
3P3W3A(3P4W)
Q
1,
Q
2,
Q
3,
Q
sum
=
Q
1
+Q
2
+Q
3