beautypg.com

Safety warnings – KYORITSU 6315 Instruction Manual User Manual

Page 10

background image

KEW6315

Safety warnings

KEW6315

- 8 -

Safety warnings

This instrument has been designed, manufactured and tested according to IEC 61010-1: Safety requirements
for Electronic Measuring apparatus, and delivered in the best condition after passing quality control tests.

This instruction manual contains warnings and safety procedures which have to be observed by the user
to ensure safe operation of the instrument and to maintain it in safe condition. Therefore, read through
these operating instructions before starting to use the instrument.

WARNING

- For about Instruction manual -
● Read through and understand the instructions contained in this manual before using the instrument.
● Keep the manual at hand to enable quick reference whenever necessary.
● The instrument is to be used only in its intended applications.
● Understand and follow all the safety instructions contained in the manual.
● Read the enclosed Quick manual after reading this instruction manual.
● As to the Clamp sensor use, refer to the instruction manual supplied with the sensor.
It is essential that the above instructions are adhered to. Failure to follow the above instructions may cause

injury, instrument damage and/or damage to equipment under test. Kyoritsu assumes no responsibility for

damage and injury caused by misuse or not following the instructions in the manual.

The symbol indicated on the instrument, means that the user must refer to the related parts in the manual for
safe operation of the instrument. It is essential to read the instructions wherever the symbol appears in the manual.

DANGER

: is reserved for conditions and actions that are likely to cause serious or fatal injury.

WARNING

: is reserved for conditions and actions that can cause serious or fatal injury.

CAUTION

: is reserved for conditions and actions that can cause injury or instrument damage.

Measurement Category

To ensure safe operation of measuring instruments, IEC 61010 establishes safety standards for various
electrical environments, categorized as O to CAT.IV, and called measurement categories. Higher-numbered
categories correspond to electrical environments with greater momentary energy, so a measuring instrument
designed for CAT.III environments can endure greater momentary energy than one designed for CAT.II.

O

: Circuits which are not directly connected to the mains power supply.

CAT.II

: Electrical circuits of equipment connected to an AC electrical outlet by a power cord.

CAT.III

: Primary electrical circuits of the equipment connected directly to the distribution panel,

and feeders from the distribution panel to outlets.

CAT.IV

: The circuit from the service drop to the service entrance, and to the power meter and

primary overcurrent protection device (distribution panel).