Predicted value (dem, Present value, measured demand value (σdem), Σdem – KYORITSU 6315 Instruction Manual User Manual
Page 145: Wpsum (+wssum), Dem σ

Items measured at Demand measurement
KEW6315
- 143 -
KEW6315
Items measured at Demand measurement
Target value (DEM
Target
)
Displayed digit
4-digit
Unit
m, k, M, G, T
Display range
0.000mW(VA) - 999.9TW(VA)
*according to the selected values
Predicted value (DEM
Guess
)
Displayed digit
6-digit
Unit
m, k, M, G, T (depending on DEM
Target value
)
Display range
0.00000mW(VA) - 99999.9TW(VA)
* Decimal point is dependent on the DEM
Target.
* “OL” is displayed when the display area is exceeded.
Equation
time
Elapsed
erval
Demand
DEM
int
DEM
Guess
×
Σ
=
Present value, Measured demand value (ΣDEM)
Displayed digit
6-digit , Unit: m, k, M, G, T (depending on DEM
Target value)
Unit
m, k, M, G, T (depending on DEM
Target value)
Display range
0.00000mW(VA) - 99999.9TW(VA)
* Decimal point is dependent on the DEM
Target.
* “OL” is displayed when the display area is exceeded.
Equation
ΣDEM=
(Integration values of "
+WPsum (+WSsum)
")
Interval
hour
1
×
Load factor
Displayed digit
6-digit
Display range
0.00 - 9999.99% * “OL” is displayed when the display area is exceeded.
Equation
Terget
DEM
DEM
Σ
Estimation
Displayed digit
6-digit
Display range
0.00 - 9999.99% * “OL” is displayed when the display area is exceeded.
Equation
Terget
Guess
DEM
DEM