3 ”demand, Showing the measured values, Instantaneous value “w – KYORITSU 6315 Instruction Manual User Manual
Page 104

KEW6315
6.3 Demand
KEW6315
- 102 -
6.3 ”Demand”
Press the
Key.
Display the screen for demand value.
Change the screens to display the demand measurement results in various forms.
Showing the measured values
Move the blue highlight to “Meas.”.
The demand is the average powers recorded over a certain period. When the estimated value exceeds the
target value during demand measurements, the warning buzzer sounds at the inspection cycles.
Items displayed on the LCD
Remaining time
(time left)
Demand interval is counted down.
DEM Target
Demand target value.
DEM Guess
Predicted demand value (average power) when preset demand interval elapses
under present load.
(Present
(Present
value)
x interval)
(Elapsed time)
* Integration and calculations are done as time
elapses.
DEM Present
Demand value (average power) within a demand interval.
“
““
“WP+ x 1 hour”
Interval
* Integration and calculations are done
as time elapses.
DEM Max
Recorded date
Max demand recorded during a measuring period is displayed. Displayed value will be
refreshed if any higher demand is detected.
Instantaneous value “W”
Press the
(W) Key to show instantaneous values on the screen. Please refer to “6.1
Instantaneous value “W”” (P. 92) in this manual for further details.