11.2 Error messages and actions
KEW6315
- 160 -
Message
Detail & Action
New sensor is detected.
Recheck the basic setting for SET
UP before measurements.
● The connected Clamp sensors are not the same ones used
during the previous test. Modify the settings of clamp sensor
directly from the “Basic setting” or press the “Detect” key.
Sensor connection is not correct.
Check the connected sensor(s).
● Appropriate current sensor may not be connected to the measure-
ment channels. Check the wiring configuration and the connected
sensor.
Out of SD card space.
Recording will be stopped.
● First, stop the recording. Confirm “Recording stopped.” message
appears, and then disappears. Backup the data file to PC or any
other medias, and then delete files or format. When using another
SD card, it should be formatted on KEW6315, not on the PC.
See “To delete, transfer or format the recorded data” (
P.
82).
Out of internal memory space.
Recording will be stopped.
● First, stop the recording. Confirm “Recording stopped.” message
appears, and then disappears. Backup the data file to PC or SD
cards, and then delete files or format.
See “To delete, transfer or
format the recorded data” (
P.
82).