beautypg.com

Shows a-zero setting, Select the setting using the up/down key – GW Instek GDM-8261 User Manual User Manual

Page 88

background image

GDM-8261 User Manual

88

The diagram below shows how the total offset is

obtained.

Buffer

Hi

Total offset

A/D

Driver

ADC

Lo

ADC

offset

A/D Driver

offset

Buffer

offset

Input Hi

Terminal

Input Lo

Terminal

Applicable
Measurement
Mode, Rate and
Speed settings

Mode

Rate

Accuracy Speed

Quick Speed

DCV,
DCI,
4W/2W

S

M

F

TC, RTD  (No Speed setting)

Panel operation

1. Press the Shift key followed by

the 2nd (Menu) key. Press the

right key twice. The ADC

setting menu appears.

/ EXIT

SHIFT

2ND

MENU

TRIG

TRIG

2. Press the Down key twice. The

A-Zero setting appears.

1st display

Shows A-Zero setting

3. Select the setting using the

Up/Down key.

4. Press the Enter key to confirm

your selection. Press the Exit

key to go back to the default

display.

AUTO

ENTER

/ EXIT

SHIFT