REMOTE CONTROL
189
[SENSe:]TEMPerature:FRTD:RESolution
Sets the 4W RTD measurement resolution. The resolution depends on the
rate and range settings.
Parameter: Resolution(
Example: SENS:TEMP:FRTD:RES MAX
Sets the 4W RTD resolution to the maximum.
[SENSe:]TEMPerature:FRTD:RESolution?
Returns the 4W RTD measurement resolution.
Parameter: [None] | [MIN | MAX]
[SENSe:]TEMPerature:RTD:RESolution
Sets the 2W RTD measurement resolution. The resolution depends on the
rate and range settings.
Parameter: Resolution(
Example: SENS:TEMP:RTD:RES MAX
Sets the 2W RTD resolution to the maximum.
[SENSe:]TEMPerature:RTD:RESolution?
Returns the 2W RTD measurement resolution.
Parameter: [None] | [MIN | MAX]
[SENSe:]VOLTage:DC:NPLCycles
Sets the integration time for DC Voltage measurements in PLCs (power line
cycles). Where one PLC is equal to 16.6 milliseconds. For any
parameter, the DMM will automatically set the PLC to the closest
acceptable PLC value (0.025, 0.1, 0.25, 1, 2, 12).
Parameter: NPLCycles(
Example: SENS:VOLT:DC:NPLC 12
Sets the integration time to 12 PLCs for DC Voltage.
[SENSe:]VOLTage:DC:NPLCycles?
Returns the integration time for DC Voltage measurement in PLCs (power
line cycles). Where one PLC is equal to 16.6 milliseconds.
Return parameter: 0.025, 0.1, 0.25, 1, 2, 12
[SENSe:]CURRent:DC:NPLCycles
Sets the integration time for DC Current measurements in PLCs (power line
cycles). Where one PLC is equal to 16.6 milliseconds. For any
parameter, the DMM will automatically set the PLC to the closest
acceptable PLC value (0.025, 0.1, 0.25, 1, 2, 12).
Parameter: NPLCycles(
Example: SENS:CURR:DC:NPLC 2
Sets the integration time to 2 PLCs for DC Current.