GW Instek GDM-8261 User Manual User Manual
Page 188
GDM-8261 User Manual
188
[SENSe:]FRESistance:RESolution
Sets the 4W Resistance measurement resolution. The resolution depends
on the rate and range settings.
Parameter: Resolution(
Example: SENS:FRES:RES 0.01
Sets the 4W Resistance resolution to 0.01
[SENSe:]FRESistance:RESolution?
Returns the 4W Resistance resolution.
Parameter: [None] | [MIN | MAX]
[SENSe:]CONTinuity:RESolution
Sets the Continuity measurement resolution. The resolution depends on the
rate and range settings.
Parameter: Resolution(
Example: SENS:CONT:RES 0.001
Sets the Continuity resolution to 0.001
[SENSe:]CONTinuity:RESolution?
Returns the Continuity measurement resolution.
Parameter: [None] | [MIN | MAX]
[SENSe:]DIODe:RESolution
Sets the Diode measurement resolution. The resolution depends on the
rate and range settings.
Parameter: Resolution(
Example:SENS:DIOD:RES 0.1e-4
Sets the Diode resolution to 0.00001
[SENSe:]DIODe:RESolution?
Returns the Diode measurement resolution.
Parameter: [None] | [MIN | MAX]
[SENSe:]TEMPerature:TCOuple:RESolution
Sets the thermocouple (T-CUP) measurement resolution. The resolution
depends on the rate and range settings.
Parameter: Resolution(
Example: SENS:TEMP:TCO:RES MAX
Sets the thermocouple resolution to the maximum.
[SENSe:]TEMPerature:TCOuple:RESolution?
Returns the thermocouple measurement resolution.
Parameter: [None] | [MIN | MAX]