CALCulate:NULL:OFFSet

Sets the reference value for the relative function. This command is
analogous to the CALCulate:REL:REFerence command.
Parameter: | MIN | MAX
Example: CALC:NULL:OFFS MAX
Sets the reference value to the maximum allowed.

CALCulate:NULL:OFFSet?

Returns the reference value from the relative function. This query is
analogous to the CALCulate:REL:REFerence? query.

TRIGger Commands

READ?

Returns 1

st

and 2

nd

display value. The Read query will not return the unit or

count number of the reading.

VAL1?

Returns the 1

st

display reading in the unit format specified in the

Configuration menu (Return Format, page 133) or from the
SYSTem:OUTPut:FORMat command (page 196).
Example: SAMP:COUN 100

VAL1?
>+0.333E-4,V DC
>+0.389E-4,V DC
> etc, for 100 counts.

Queries 100 counts of stored samples from the 1

st

display.

VAL2?

Returns the 2

nd

display reading in the unit format specified in the

Configuration menu (Return Format, page 133) or from the
SYSTem:OUTPut:FORMat command (page 196).
Example: SAMP:COUN 100

VAL2?
>+0.345E-4,V DC
>+0.391E-4,V DC
> etc, for 100 counts.

Queries 100 counts of stored samples from the 2

nd

display.

TRIGger:SOURce

Selects the trigger source.
Parameter: INT | EXT
Example: TRIG:SOUR INT
Sets the trigger source as internal.

TRIGger:SOURce?

Returns current trigger source.