*ESR?

Returns and clears SESR (Standard Event Status Register).
Example: *ESR?
>198
Returns 198. SESR=11000110

*IDN?

Returns the manufacturer, model No., serial number and system version
number.
Example: *IDN?
>GWInstek,GDM8261,00000000,1.0

*OPC?

“1” is placed in the output queue when all the pending operations are
completed.

*OPC

Sets operation complete bit (bit0) in SERS (Standard Event Status Register)
when all pending operations are completed.

*PSC?

Returns power On clear status.
Return parameter: (0 | 1) 0=clear, 1=don’t clear

*PSC

Clears power On status.
Parameter: (0|1) 0=clear, 1=don’t clear

*RST

Recalls default panel setup (reset the device).

*SRE?

Returns the SRER (Service Request Enable Register) contents.

*SRE

Sets SRER contents.
Parameter: (0~255)
Example: *SRE 7
Sets the SRER to 00000111.

*STB?

Returns the SBR (Status Byte Register) contents.
Example:*STB?
>81
Returns the contents of the SBR as 01010001.

*TRG

Manually triggers the GDM-8261.