beautypg.com

Relative value measurement, Indicates relative value measurement, Shows the measurement range – GW Instek GDM-8261 User Manual User Manual

Page 62: Shows the reference (rel)value at full scale

background image

GDM-8261 User Manual

62

Relative Value Measurement

Applicable to

ACV

DCV

ACI

DCI

2/4W

Hz/P

TEMP

Background

Relative measurement stores a value, typically the data at

the moment, as the reference. The following measurement

is shown as the delta between the reference. The reference
value will be cleared upon exit.

1. Activate
Relative
measurement

Press the REL key. The

measurement reading at the
moment becomes the reference

value.

REL

2. Relative
measurement
display appears

V

REL

S

AC

REL

Indicates Relative value measurement

2nd display

Shows the measurement range.

1st display

Shows the delta between the current

measurement data and the reference value

View reference
(REL) value

Press the 2nd key to view the
reference (REL) value.

2ND

Reference
(REL)
measurement
display appears

V

2nd display

Indicates that the reference (REL) value is
displayed on the 1st display

1st display

Shows the reference (REL)value at full

scale

Manually set
the reference
value

1. To set the reference (REL)

value manually, press the Shift

key followed by the REL key.
The setting appears.

/ EXIT

SHIFT

REL

REL#