SAFETY INSTRUCTIONS

7

Safety Guidelines

General Guideline

CAUTION

Make sure that the voltage input level does not exceed

DC1000V/AC750V.

Make sure the current input level does not exceed 10A.

Do not place any heavy object on the GDM-8261.

Avoid severe impact or rough handling that can lead to

damaging the GDM-8261.

Do not discharge static electricity to the GDM-8261.

Use only mating connectors, not bare wires, for the
terminals.

Do not block or obstruct the cooling fan vent opening.

Do not perform measurement at the source of a

low-voltage installation or at building installations
(Note below).

Do not disassemble the GDM-8261 unless you are

qualified as service personnel.

(Note) EN 61010-1:2010 specifies the measurement categories and

their requirements as follows. The GDM-8261 falls under category
I or II.

Measurement category IV is for measurement performed at the

source of low-voltage installation.

Measurement category III is for measurement performed in the
building installation.

Measurement category II is for measurement performed on the

circuits directly connected to the low voltage installation.

Measurement category I is for measurements performed on circuits

not directly connected to Mains.

Power Supply

WARNING

AC Input voltage: 100/120/220/240 V AC ±10%,
45Hz to 66Hz / 360Hz to 440Hz

The power supply voltage should not fluctuate more

than 10%.

Connect the protective grounding conductor of the
AC power cord to an earth ground, to avoid electrical

shock.