BASIC MEASUREMENT

31

Crest Factor Table

Background

Crest factor is the ratio of the peak signal amplitude to the

RMS value of the signal. It determines the accuracy of AC

measurement.

If the crest factor is less than 3.0, voltage measurement will
not result in error due to dynamic range limitations at full

scale.

If the crest factor is more than 3.0, it usually indicates an

abnormal waveform as seen from the below table.

Waveform

Shape

Crest factor

Square wave

1.0

Sine wave

1.414

Triangle
sawtooth

1.732

Mixed
frequencies

1.414 ~ 2.0

SCR output
100% ~ 10%

1.414 ~ 3.0

White noise

3.0 ~ 4.0

AC Coupled
pulse train

>3.0

Spike

>9.0