Measuring characteristics, Operating characteristics [8 – GW Instek GDM-8261 User Manual User Manual

Page 217

background image

APPENDIX

217

Measuring Characteristics

True RMS AC
Voltage

Measurement
Method:

AC-coupled True RMS – measures the ac
component of input with up to 400 Vdc of
bias on any range.

Crest Factor

Maximum 5:1 at full scale

AC Filter
Bandwidth

Slow

3 Hz – 300 kHz

Medium

20 Hz – 300 kHz

Fast

200 Hz – 300 kHz

Input
Impedance:

1 MΩ ± 2%, in parallel with 100 pF

Input
Protection:

750 Vrms on all ranges

True RMS AC
Current

Range

Shunt

Burden Voltage

1mA

100Ω

<0.15V

10mA

5Ω

<0.07V

100mA

5Ω

<0.7V

1A

0.1Ω

<0.8V

10A

10mΩ

<0.5V

Input
Protection:

Externally accessible 1.25A, 250 V fuse

Internal 12A, 250 V fuse

Operating Characteristics

[8]

Function

Rate

Digits Readings/s

AC Bandwidth

ACV,ACI

Slow

6 ½

1.2

(sec/reading)

3 Hz – 300 kHz

Medium 5 ½

3.38

20 Hz – 300 kHz

Fast

4 ½

30

200 Hz – 300 kHz

[ 1 ] Specifications are for 1-hour warm-up at 6 1⁄2 digits, Slow ac filter,
sinewave input.
[ 2 ] Relative to calibration standards.
[ 3 ] 20% overrange on all ranges, except 750 Vac, 10A range.
[ 4 ] Specifications are for sinewave input >5% of range. For inputs from
1% to 5% of range and <50 kHz, add 0.1% of range additional error. For 50
kHz to 100 kHz, add 0.13% of range.
[ 5 ] 750 Vac range limited to 100 kHz
[ 6 ] Typically 30% of reading error at 1 MHz.
[ 7 ] For frequencies below 100 Hz, slow AC filter specified for sinewave
input only.
[ 8 ] Additional settling delay required when input dc level varies.
[ 9 ] 0˚C~18˚C, 28˚C~55˚C