Filter count

Filter count defines the number of samples to be
averaged per reading. More samples offer low noise but

long delay. Less samples offer high noise but short delay.

Range

2 ~ 100

Filter Setting

Turn on Filter

1. Press the Shift key followed by

the MX/MN (Filter) key.

/ EXIT

SHIFT

MX/MN

FILTER

1st display

Shows the filter count

2nd display

Shows the filter type (flashing)

2. Select the filter type using the

Up/Down key.

3. Move the cursor to filter count

using the Left/Right key.

Change the value using the

Up/Down key.

HOLD

TRIG

4. Press the Enter key to confirm

editing. The Filter indicator
appears on the display.

AUTO

ENTER

V

m

S

DC

FILT

FILT

Indicates manual Filter setting