GDM-8261 User Manual

192

CALCulate:LIMit:UPPer?

Returns the upper limit of the compare function.

CALCulate:DB:REFerence

Sets the reference value for the dB function.
Parameter: | MIN | MAX
Example: CALC:DB:REF MAX
Sets the reference voltage for dB measurements to the maximum allowed.

CALCulate:DB:REFerence?

Returns the reference voltage from the dB function.

CALCulate:DBM:REFerence

Sets the resistance value for the dBm function.
Parameter: | MIN | MAX
Example: CALC:DBM:REF MAX
Sets the resistance value for dBm measurements to the maximum allowed.

CALCulate:DBM:REFerence?

Returns the resistance value from the dBm function.

CALCulate:STORe:COUNt

Set the number of measurement counts that are recorded with the Store
measurement function.
Parameter: (2 ~ 9999) | MIN | MAX
Example: CALC:STOR:COUN 1000
Sets the number of counts to be recorded as 1000.

CALCulate:STORe:COUNt?

Returns the number of counts that are recorded with the Store
measurement function.
Parameter: [None] | MIN | MAX

CALCulate:AVERage:COUNt

Sets the total number of statistic counts.
Parameter: (0, 2~100000) 0=continuous count, 2~100000=count
Example: CALC:AVER:COUN 0
Sets the count to continuous.

CALCulate:AVERage:COUNt?

Returns the total number of recorded counts. The setting commands for
this query are: CALCulate:STORe:COUNt, ROUTe:COUNt and
CALCulate:AVERage:COUNt.
Parameter: None | (0~2) 0=Store, 1=Scan, 2=Stats
Example: CALC:AVER:COUN? 0

>+0010

Returns the total number of counts set for the Store function (10 counts).