ADVANCED MEASUREMENT

57

Refresh Rate

Function

S

M

F

(Readings/s)

Continuity / Diode

100

200

300

DCV/DCI/100Ω~
100MΩ (Accurate)

5

60

240

DCV/DCI/100Ω~
100MΩ (Quick)

30

600

2400

ACV/ACI

1.2

(sec/reading)

3.38

30

Frequency/Period

1

10

100

Selection steps

1. Press the Shift key followed by

the AUTO (RATE) key. The

refresh rate switches to the next.

/ EXIT

SHIFT

AUTO

RATE

2. The refresh rate indicator shows

the current status.

S

M

F

S

Reading Indicator

Background

The reading indicator next to the 1st display flashes

according to the refresh rate when the captured data is

updated on the display.

V

m

When no data is
captured

When there is no captured data, the reading indicator

flashes once every two seconds (slower than the normal

refresh rate), indicating the DMM is in the waiting mode.

Common Attribute: Manual/Automatic Triggering

Automatic
triggering
(default)

The GDM-8261triggers according to the refresh rate. See

the previous page for refresh rate setting details.