Calculate commands – GW Instek GDM-8261 User Manual User Manual
Page 190
GDM-8261 User Manual
190
[SENSe:]CURRent:DC:NPLCycles?
Returns the integration time for DC Current measurement in PLCs (power
line cycles). Where one PLC is equal to 16.6 milliseconds.
Return parameter: 0.025, 0.1, 0.25, 1, 2, 12
[SENSe:]RESistance:NPLCycles
Sets the integration time for 2W resistance measurements in PLCs (power
line cycles). Where one PLC is equal to 16.6 milliseconds. For any
parameter, the DMM will automatically set the PLC to the closest
acceptable PLC value (0.025, 0.1, 0.25, 1, 2, 12).
Parameter: NPLCycles(
Example: SENS:RES:NPLC MIN
Sets the integration time to 0.025 PLCs for 2W resistance measurements.
[SENSe:]RESistance:NPLCycles?
Returns the integration time for 2W resistance measurements in PLCs
(power line cycles). Where one PLC is equal to 16.6 milliseconds.
Return parameter: 0.025, 0.1, 0.25, 1, 2, 12
[SENSe:]FRESistance:NPLCycles
Sets the integration time for 4W resistance measurements in PLCs (power
line cycles). Where one PLC is equal to 16.6 milliseconds. For any
parameter, the DMM will automatically set the PLC to the closest
acceptable PLC value (0.025, 0.1, 0.25, 1, 2, 12).
Parameter: NPLCycles(
Example: SENS:FRES:NPLC MAX
Sets the integration time to the maximum for 4W resistance measurements.
[SENSe:]FRESistance:NPLCycles?
Returns the integration time for 4W resistance measurements in PLCs
(power line cycles). Where one PLC is equal to 16.6 milliseconds.
Return parameter: 0.025, 0.1, 0.25, 1, 2, 12
CALCulate Commands
CALCulate:FUNCtion
Sets the Advanced function.
Parameter: OFF | MIN | MAX | HOLD | REL | COMP | DB | DBM | STORE |
AVER | MXB | INV | REF
Example: CALC:FUNC REL
Sets the Advanced function to REL (relative)
CALCulate:FUNCtion?
Returns the current Advanced function.