GW Instek GDM-8261 User Manual User Manual
Page 180

GDM-8261 User Manual
180
[SENSe:]AVERage:TCONtrol
Selects the digital filter.
Parameter: MOV | REP
Example: SENS:AVER:TCON MOV
Sets the digital filter to the Moving filter.
[SENSe:]AVERage:TCONtrol?
Returns the current digital filter type.
Return parameter: MOV (moving), REP (repeating)
[SENSe:]AVERage:COUNt
Sets the digital filter count.
Parameter:
Example: SENS:AVER:COUN 100
Sets the digital filter count number to 100.
[SENSe:]AVERage:COUNt?
Returns the digital filter count.
Return parameter:
[SENSe:]AVERage:STATe
Turns the digital filter On/Off.
Parameter: ON | OFF
Example: SENS:AVER:STAT ON
Turns the digital filter on.
[SENSe:]AVERage:STATe?
Returns the state of the digital filter (on or off).
Return parameter: 0|1, 0=OFF, 1=ON
[SENSe:]FREQuency:APERture
Sets the aperture time (gate time) for the frequency function (0.01=F,
0.1=M, 1=S).
Parameter: (0.01 | 0.1 | 1)
Example: SENS:FREQ:APER 0.01
Sets the gate time to 0.01 seconds.
[SENSe:]FREQuency:APERture?
Returns aperture time (gate time) for the frequency function.
[SENSe:]PERiod:APERture
Sets the aperture time (gate time) for the period function(0.01=F, 0.1=M,
1=S).
Parameter:
Example: SENS:PER:APER 0.1
Sets the gate time to 0.1 seconds for the period function.