beautypg.com

Select thermocouple type – GW Instek GDM-8261 User Manual User Manual

Page 44

background image

GDM-8261 User Manual

44

Range

RTD: -200°C ~ +600°C (sensor dependent)
Thermocouple: -210°C ~ +1820°C (sensor dependent)

Select Thermocouple Type

Background

The GDM-8261accepts thermocouple inputs and
calculates the temperature from the voltage difference of

two dissimilar metals. Thermocouple type and reference

junction temperature are also considered.

Parameter

Thermocouple

Range

Resolution

E

-200 to +1000°C 0.002 °C

J

-210 to +1200˚C 0.002 °C

T

-200 to +400°C

0.002 °C

K

-200 to +1372˚C 0.002 °C

N

-200 to +1300˚C 0.003 °C

R

-50 to +1768˚C

0.01 °C

S

-50 to +1768˚C

0.01 °C

B

+350 to +1820˚C 0.01 °C

1. Open sensor
selection menu

Press the Shift key, then the TEMP

(Sensor) key. The sensor selection

menu appears on the display.

/ EXIT

SHIFT

TEMP

SENSOR

2. Select sensor
type

Press the Left and Right arrow keys

and select T-CUP (thermocouple).

HOLD

TRIG

3. Select sensor

Press the Down key twice. The

sensor selection menu appears on
the display.