beautypg.com

Temperature characteristics, Display in °c, °f, exclusive of probe errors.) – GW Instek GDM-8261 User Manual User Manual

Page 219

background image

APPENDIX

219

Temperature Characteristics

(Display in °C, °F, Exclusive of probe errors.)

RTD (Accuracy based on PT100):

(100Ω platinum [PT100], D100, F100,PT385, PT3916, or user type)

[1]

Range

Resolution

1 Year (23°C ±5°C)

Temperature Coefficient
0°-18°C & 28°-55°C

-200°C~ -100°C

0.001°C

0.09°C

0.004 °C / °C

-100°C~-20°C

0.001°C

0.08°C

0.005 °C / °C

-20°C~20°C

0.001°C

0.06°C

0.005 °C / °C

20°C~100 °C

0.001°C

0.08°C

0.005 °C / °C

100°C~300 °C

0.001°C

0.12°C

0.007 °C / °C

300°C~600 °C

0.001°C

0.22°C

0.009 °C / °C

Thermocouples (Accuracy based on ITS-90)

[1]

:

Type Range

Resolution

90 Day/1 Year
(23°C±5°C)*

Temperature Coefficient
0°-18°C & 28°-55°C

E

-200 to +1000°C

0.002 °C

0.2 °C

0.03 °C / °C

J

-210 to +1200˚C

0.002 °C

0.2 °C

0.03 °C / °C

T

-200 to +400°C

0.002 °C

0.3 °C

0.04 °C / °C

K

-200 to +1372˚C

0.002 °C

0.3 °C

0.04 °C / °C

N

-200 to +1300˚C

0.003 °C

0.4 °C

0.05 °C / °C

R

-50 to +1768˚C

0.01 °C

1 °C

0.14 °C / °C

S

-50 to +1768˚C

0.01 °C

1 °C

0.14 °C / °C

B

+350 to +1820˚C

0.01 °C

1 °C

0.14 °C / °C

[ 1 ] Specifications do not include probe accuracy
*Relative to simulated junction