Secondary display: configure2 commands – GW Instek GDM-8261 User Manual User Manual
Page 171

REMOTE CONTROL
171
CONFigure:TEMPerature:RTD
Sets the measurement mode to 2W RTD measurement mode on the first 
display. Sets the sensor type. 
Parameter: [None] | [Type(PT100 | D100 | F100 | PT385 | PT3916 | USER)] 
Example: CONF:TEMP:RTD PT100 
Sets the sensor type to PT100 and sets the measurement mode to 2W RTD 
CONFigure:FUNCtion?
Returns the current function on 1st display. 
Return parameter: VOLT, VOLT:AC, CURR, CURR:AC, RES, FRES, FREQ, 
PER, TEMP:RTD, TEMP:FRTD, TEMP:TCO, DIOD, CONT 
CONFigure:RANGe?
Returns the current range on 1
st
display.
Return Parameter: 
DCV: 0 .1(100mV), 1(1V), 10(10V), 100(100V), 1000(1000V) 
ACV: 0.1(100mV), 1(1V), 10(10V), 100(100V), 750(750V) 
ACI: 0.001 (1mA), 0.01(10mA), 0.1(100mA), 1(1A), 10(10A) 
DCI: 0.0001 (100μA), 0.001 (1mA), 0.01(10mA), 0.1(100mA), 1(1A), 
10(10A) 
RES: 10E+1(100Ω) 10E+2(1kΩ), 10E+3(10kΩ), 10E+4 (100kΩ), 
10E+5(1MΩ), 10E+6(10MΩ), 10E+7(100MΩ) 
CONFigure:AUTO
Sets Auto-Range on or off on the first display. 
Parameter: ON | OFF 
Example: CONF:AUTO ON 
CONFigure:AUTO?
Returns the Auto-Range status of the function on the 1
st
display.
Return Parameter: 0|1, 1=Auto range, 0=Manual range
Secondary Display: CONFigure2 Commands
CONFigure2:VOLTage:DC
Sets measurement to DC Voltage on the second display and specifies 
range/resolution. 
Parameter: [None] | [Range(
MIN | MAX | DEF)] 
Example: CONF2:VOLT:DC 1,MAX 
Sets the voltage range to 1 volt and the resolution to the maximum. 
