REMOTE CONTROL

171

CONFigure:TEMPerature:RTD

Sets the measurement mode to 2W RTD measurement mode on the first
display. Sets the sensor type.
Parameter: [None] | [Type(PT100 | D100 | F100 | PT385 | PT3916 | USER)]
Example: CONF:TEMP:RTD PT100
Sets the sensor type to PT100 and sets the measurement mode to 2W RTD

CONFigure:FUNCtion?

Returns the current function on 1st display.
Return parameter: VOLT, VOLT:AC, CURR, CURR:AC, RES, FRES, FREQ,
PER, TEMP:RTD, TEMP:FRTD, TEMP:TCO, DIOD, CONT

CONFigure:RANGe?

Returns the current range on 1

st

display.

Return Parameter:
DCV: 0 .1(100mV), 1(1V), 10(10V), 100(100V), 1000(1000V)
ACV: 0.1(100mV), 1(1V), 10(10V), 100(100V), 750(750V)
ACI: 0.001 (1mA), 0.01(10mA), 0.1(100mA), 1(1A), 10(10A)
DCI: 0.0001 (100μA), 0.001 (1mA), 0.01(10mA), 0.1(100mA), 1(1A),
10(10A)
RES: 10E+1(100Ω) 10E+2(1kΩ), 10E+3(10kΩ), 10E+4 (100kΩ),
10E+5(1MΩ), 10E+6(10MΩ), 10E+7(100MΩ)

CONFigure:AUTO

Sets Auto-Range on or off on the first display.
Parameter: ON | OFF
Example: CONF:AUTO ON

CONFigure:AUTO?

Returns the Auto-Range status of the function on the 1

st

display.

Return Parameter: 0|1, 1=Auto range, 0=Manual range

Secondary Display: CONFigure2 Commands

CONFigure2:VOLTage:DC

Sets measurement to DC Voltage on the second display and specifies
range/resolution.
Parameter: [None] | [Range( | MIN | MAX | DEF),Resolution( |
MIN | MAX | DEF)]
Example: CONF2:VOLT:DC 1,MAX
Sets the voltage range to 1 volt and the resolution to the maximum.