beautypg.com

Select resistance range – GW Instek GDM-8261 User Manual User Manual

Page 35

background image

BASIC MEASUREMENT

35

2. 2W/4W
resistance mode
display appears

W

S

2W

k

AUTO

2W or 4W + Ω Indicates 2W or 4W Resistance mode

AUTO

Indicates Automatic range selection

1K

2nd display shows the Resistance range

3. Connect the
test lead and
measure

Connect the test lead. For 2-wire resistance, use the Ω (V)
and the COM port. For 4-wire resistance, use the Ω (V)

and the COM port, plus the 4W sense, and LO port for

sensing. The display updates the reading.

2W connection

INPUT

V

W

COM

4W connection

W

4W

INPUT

SENSE

HI

LO

V

W

COM

Select Resistance Range

Auto range

To turn the automatic range

selection On/Off, press the AUTO

key.

AUTO

Manual range

Press the Up or the Down key to

select the range. AUTO indicator

turns Off automatically. If the range
is unknown, select the highest range.

Selection list

Range

Resolution Full scale @ slow rate

100Ω

0.1μΩ

119.9999Ω

1kΩ

1μΩ

1.199999kΩ

10kΩ

10μΩ

11.99999kΩ

100kΩ

100μΩ

119.9999kΩ

1MΩ

1.199999MΩ