Application: compare measurement, Eom pulse width timing, Press the shift key, then the hold (comp) key – GW Instek GDM-8261 User Manual User Manual
Page 122: Shows the high limit value, Indicates high limit setting
GDM-8261 User Manual
122
Pin6
FAIL signal Output. Activates when the
compare result is FAIL.
Pin7
HIGH Limit FAIL signal Output. Activates
when the compare result is FAIL due to
violating the HIGH Limit.
Pin8
LOW Limit FAIL signal Output. Activates
when the compare result is FAIL due to
violating the LOW Limit.
Pin9
EOM (End Of Measurement) signal Output.
Activates when compare measurement is over.
It is also available in other measurements.
EOM pulse
width timing
>10
μs (R=1kΩ)
0V
+5V
Application: Compare measurement
Applicable to
ACV
DCV
ACI
DCI
2/4W
Hz/P
TEMP
Background
Compare measurement checks and updates if the
measurement data stays between the upper (high) and
lower (low) limit specified.
1. Activate
Compare
measurement
Press the Shift key, then the Hold
(Comp) key.
/ EXIT
SHIFT
HOLD
COMP
2. High limit
setting
V
1st display
Shows the high limit value
2nd display
Indicates high limit setting
1. Use the Left/Right key to move
the cursor (flashing point)
between high/low setting, digits,
and decimal point.