beautypg.com

Application: compare measurement, Eom pulse width timing, Press the shift key, then the hold (comp) key – GW Instek GDM-8261 User Manual User Manual

Page 122: Shows the high limit value, Indicates high limit setting

background image

GDM-8261 User Manual

122

Pin6

FAIL signal Output. Activates when the
compare result is FAIL.

Pin7

HIGH Limit FAIL signal Output. Activates

when the compare result is FAIL due to
violating the HIGH Limit.

Pin8

LOW Limit FAIL signal Output. Activates

when the compare result is FAIL due to
violating the LOW Limit.

Pin9

EOM (End Of Measurement) signal Output.

Activates when compare measurement is over.
It is also available in other measurements.

EOM pulse

width timing

>10

μs (R=1kΩ)

0V

+5V

Application: Compare measurement

Applicable to

ACV

DCV

ACI

DCI

2/4W

Hz/P

TEMP

Background

Compare measurement checks and updates if the

measurement data stays between the upper (high) and

lower (low) limit specified.

1. Activate
Compare
measurement

Press the Shift key, then the Hold

(Comp) key.

/ EXIT

SHIFT

HOLD

COMP

2. High limit
setting

V

1st display

Shows the high limit value

2nd display

Indicates high limit setting

1. Use the Left/Right key to move

the cursor (flashing point)

between high/low setting, digits,

and decimal point.

HOLD

TRIG