5 reference junction compensation accuracy test – Yokogawa PC-Based MX100 User Manual
Page 85
2-56
SM MX100-01E
2.12.5 Reference Junction Compensation Accuracy Test
Always use terminal covers.
Procedure
1.
Allow a warm-up of 30 minutes or more.
2.
Measure the 0°C standardized temperature tank at the range for Type-T
thermocouples.
3.
Measure at all channels, and confirm that the result lies within 0°C ± 0.5°C.
2.12 Tests of the 10-CH Screw Terminal Block (772061)