6 pulse duty accuracy, Pwm outupt settings Page 77
Page 77
2-48
SM MX100-01E
2.10.6 Pulse Duty Accuracy
Wire the units in the same manner as in the “Pulse Interval Accuracy Test,” enter all
settings (PWM output settings), then confirm that the criteria are met.
The wiring diagram for the pulse interval accuracy test shows an example of wiring CH5.
Actual measurment should be performed on all channels.
Use a counter trigger level of 2.5 V.
PWM Outupt Settings
Connect to the PC, then enter the following settings using the MX100 standard software.
When using the MX100/DARWIN API for settings, enter settings that are equivalent to
the following.
Item
Setting
Remarks
Enable/disable
Enabled
Action
User output
Span
Need not specify
Output upon power ON
Preset value
Sets the output at power ON as a
preset value.
Output upon error
Preset value
Sets the output upon error as a
preset value.
Preset value
0.000 %
Sets 0% duty.
Resolution
Depends on the test
Pulse interval
Depends on the test
Other settings
Not required
Referring to the MW100 user’s manual for browser operation, or by using communication
commands, set up the MW100 or equivalent according to the table below.
Item
Setting
Remarks
Output type
PWM
Output method
COM
Command output
Output range
Depends on the test
Span
Need not specify
Pulse interval
Depends on the test
Preset value
0.000 %
Sets 0% duty.
Reference channel
Not required
Output Operation Setting
Item
Setting
Power ON
PRESET
Error occurence
PRESET
Criteria (for the MX100 and MW100)
If entering the pulse duty and PWM settings according to the table below, confirm that
the output pulse interval of the specified channels satisfies the criteria.
(Set for each channel. Channels not under test output 0%).
Pulse interval
PWM Output Setting Criteria
1 ms
1.000 %
Duty: 1 % ±0.15 % (Pulse interval: 10s ±1.5 s)
50.000 %
Duty: 50 % ±0.15 % (Pulse interval: 500 s ±1.55s)
99.000 %
Duty: 99 % ±0.15 % (Pulse interval: 990 s ±1.59s)
5 ms (1 ms ×5)
50.000 %
Duty: 50 % ±0.03 % (Pulse interval: 2.5 ms ± 1.75s)
10 ms (10 ms ×10) 1.000 %
Duty: 1 % ±0.015 % (Pulse interval: 0.1 ms ±1.51 s)
50.000 %
Duty: 50 % ±0.015 % (Pulse interval: 5 ms ±1.9 s)
99.000 %
Duty: 99 % ±0.015 % (Pulse interval: 9.9 ms ±1.9 s)
100 ms (10 ms ×10) 50.000 %
Duty: 50 % ±0.003 %
2.10 8-CH, Medium-Speed PWM Output Module