2.7.4
ID Check
This test confirms whether the module is correctly identified when module information is
obtained from the MX100, MW100, or equivalent.
(If the OFF/ON cycle is not performed on the MX100, MW100, or equivalent, the ID
information is not updated. )
2.7.5
Bridge Voltage Accuracy Test
Wire as indicated below, then measure the bridge voltage on the waveform device.
Confirm that the bridge voltage satisfies the criteria.
Measured CH
Any
Load resistance criterion
100 Ω ±1%, 0.1W or more
Bridge voltage criterion
2 V ±0.05 V
Measured waveform
Bridge voltage
0V
2.7 Tests of the 4-CH Strain Input Module