Tests, Instruments used – Yokogawa PC-Based MX100 User Manual
Page 43
2-14
SM MX100-01E
2.5 Tests of the 10-CH, Medium-Speed Universal
Input Module (MX110-UNV-H10)
Tests
2.5.1 Insulation Resistance Test
2.5.2 Withstanding Voltage Test
2.5.3 Excessive Input Test
2.5.4 ID Check
2.5.5 Measurement Accuracy Test
2.5.6 Between-Channel Error Test
2.5.7 Burnout Test
2.5.8 Reference Junction Compensation Accuracy Test
Instruments Used
Instrument
Standards, Recommended Products Etc.
MX100, MW100, or equivalent A device that can connect with and perform communications with
the MX110 10-ch Medium-Speed Universal Input Module (Ex.
MX100 + MX150 or MW100 + MX150)
MX150 Base Plate
Front TR Assembly (When with
the /NC option)
Product number: B8720BE
A device that can receive input independently on each channel
Screw Term Assembly & Cable Product number: B8720BH
Used to confirm the ID setting. If the ID is correct, a device with an
altered connector or the like can be substituted.
PC
‡
A device that can perform communication with the MX100/M\W100
or equivalent instrument
DC voltage generator
FLUKE 5520A* or equivalent
Accuracy: 0.005% of setting + 1mV
A device capable of 100 V output or higher
Rheostat
ALPHA ADR-1202
†
or equivalent
Accuracy: 0.005% of setting + 2 mΩ
A device with resolution of 1 mΩ or less
DMM
Agilent 3458A or equivalent
Voltage: 0.005% of rdg + 1mV
Resistance: 0.01% of rdg + 1 mΩ
Insulation resistance tester
DC500 V
Withstanding voltage tester
AC 1 kV–4 kV
Temperature tank standardized
at 0°C
Coper Electronics ZC-114 or equivalent
0±0.01°C
* If a DC voltage generator meeting the standards above is not available, you can ensure that
the stated voltage is output by substituting a FLUKE 9100 or equivalent and monitoring the
output voltage using a DMM. Also, due to the output impedance, some voltage generators
cannot output all stated voltages, and at 0 V, there are many cases where the above
standard is not met, so 0 V can be obtained by shorting the terminals. It is recommended to
monitor when set to DOWN.
† Measure the setting value and actual output value, then confirm that the above standards
are met.
0.2 Ω or less cannot be set on the ADR-1202, so 0 Ω can be achieved by shorting the
terminals.
‡ Ensure that the module settings and measured values can be read in.