Mx110-v4r-m06) -20, Tests, Instruments used – Yokogawa PC-Based MX100 User Manual
Page 49
2-20
SM MX100-01E
2.6 Tests of the 6-CH, Medium-Speed Four-
Wire RTD Resistance Input Module (MX110-
V4R-M06)
Tests
2.6.1 Insulation Resistance Test
2.6.2 Withstanding Voltage Test
2.6.3 Excessive Input Test
2.6.4 ID Check
2.6.5 Measurement Accuracy Test
2.6.6 Between-Channel Error Test
2.6.7 DI Contact Input Test
Instruments Used
Instrument
Standards, Recommended Products etc.
MX100, MW100, or equivalent
A device that can connect to and perform communications with
the MX110 6-CH Medium-Speed Four-wire RTD Resistance
Input Module (Ex. MX100 + MX150, or MW100 + MX150)
MX150 Base Plate
PC
‡
A device capable of Ethernet communication with the MX100,
MW100, or equivalent
DC voltage generator
FLUKE 5520A
*
or equivalent
Accuracy of ±0.005% of setting + 1 V
A device capable of output of 100 V or more
Variable resistor
ALPHA ADR-3204G
†
or equivalent
Accuracy of except for 0 Ω: ±0.006% of setting + 2 mΩ
A device with resolution of 1 mΩ or less
DMM
Agilent 3458A or equivalent
Voltage: 0.005% of rdg + 1 V
Resistance: 0.005% of rdg + 1 mΩ
Insulation resistance tester
DC500 V
Withstanding voltage tester
AC 1 kV–4 kV
0°C standardized temperature tank Coper Electronics ZC114 or equivalent
0 ±0.01°C
* If a DC voltage generator of the above standard is not available, you can ensure that the
stated voltage is output by substituting a FLUKE 9100 or equivalent and monitoring the
output voltage with a DMM. Also, due to the output impedance, some voltage generators
cannot output all stated voltages, and at 0 V, there are many cases where the above
standard is not met, so 0 V can be obtained by shorting the terminals. It is recommended to
monitor when set to DOWN.
† Measure the setting value and actual output value, then confirm that the above standards
are met.
0.2 Ω or less cannot be set on the ADR-3204G, so 0 Ω can be achieved by shorting the
terminals.
‡ Ensure that the module settings and measured values can be read in.