Yokogawa PC-Based MX100 User Manual
Page 80
2-51
SM MX100-01E
Testing
1
2
3
4
5
6
7
Item
Setting
Remarks
Output Type
PWM
Output method
COM
Command output
Output range
1 ms
Span
Need not specify
Pulse interval
10
Preset value
0.000%
Reference channel
Not required
Output Operation Setting
Item
Setting
Power ON
PRESET
Error occurence
PRESET
Criteria (for the MX100 and MW100)
Change the settings following the steps in the table below, then confirm that the results
satisfy the criteria for voltage.
Step
External Power Supply PWM Output Setting
Voltage Criterion
1
5 V
100.000 %
3.8 V or more
2
12 V
100.000 %
0.04 V or less
3
5 V
100.000 %
0.04 V or less
4
OFF
100.000 %
5
5 V
100.000 %
3.8 V or more