9 tests of the 8-ch analog output module, Tests of the 8-ch analog output module -38, Test instruments – Yokogawa PC-Based MX100 User Manual
Page 67: Tests, 1 insulation resistance test
2-38
SM MX100-01E
2.9 Tests of the 8-CH Analog Output Module
Test Instruments
Instrument
Standards, Recommended Products etc.
MX100, MW100, or equivalent
MX100 or MW100 main module
MX150 base plate
PC
That the software required for the test is installed
External power supply
24 V ±10% 0.5A or more
DMM
Accuracy Voltage ±(0.01% of rdg. +50 V) or less
Current ±(0.01% of rdg. +1 A) or less Agilent 3458A
Fixed resistance
600 Ω ±5% 0.5W or more
Insulation resistance tester
DC500 V TOS9201 (Kikusui)
Withstanding voltage tester
AC 2 kV–4 kV TOS9201 (Kikusui)
Ethernet cable
Tests
2.9.1 Insulating Resistance Test
2.9.2 Withstand Voltage Test
2.9.3 ID Check
2.9.4 Power ON Output Check
2.9.5 Output Accuracy Test
2.9.1
Insulation Resistance Test
Install the module on the MX100, MW100, or equivalent before performing this test.
Perform the withstanding voltage test before performing this test.
Wire the devices as in the figure below, then confirm that the criteria are met.
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
V+
-
I+
V+
-
I+
V+
-
I+
V+
-
I+
V+
-
I+
V+
-
I+
V+
-
I+
V+
-
I+
+
-
+
-
Insulation
wishstanding
voltage tester
CH1 to CH4
CH5 to CH8
To the earth terminal on
the MX100, MW100, or
equivalent device
Pins 13 on each output connector are not connected internally, therefore you can choose
whether or not to make the connections shown in the wiring diagram.
In order to ensure continuity from the MX100, MW100, or equivalent’s earth to the heat
sink of the module under test, you must fasten the module with screws after installing it
onto the MX150 base plate.
Terminals Measured
Criteria
Output terminal to earth terminal on the MX100,
MW100, or equivalent
20 MΩ or more at DC 500 V