beautypg.com

3 id check, 4 power on output check – Yokogawa PC-Based MX100 User Manual

Page 68

background image

2-39

SM MX100-01E

Testing

1

2

3

4

5

6

7

2.9.2

Withstanding Voltage Test

Install the module under test in the MX110, MW100, or equivalent before performing this
test.
When applying voltages, ensure that the rising and falling times are both at least five
seconds.
Wire the instrument in the same manner as in the insulation resistance test, then confirm
that the results satisfy the criteria below.

Terminals Measured

Criteria

Output terminal to earth terminal on the MX100,

MW100, or equivalent

AC2300 Vrms for 1 minute, with current

leakage of 2 mA or less

2.9.3

ID Check

This test confirms whether the module is correctly identified when module information is
obtained from the MX100, MW100, or equivalent.
(If the OFF/ON cycle is not performed on the MX100, MW100, or equivalent, the ID
information is not updated.)

2.9.4

Power ON Output Check

Wire the units as shown in the connection diagram, then perform the power ON output
check. The figure below shows an example of wiring for measurement on CH5. Connect
other channel in the same manner and perform measurement.

Precautions to Be Taken When Connecting Cables

▪ For current measurements, connect 600 Ω series resistance as shown in the figure

below before taking measurements.

When using a DMM whose current measuring range is of a large internal resistance,

adjust the external resistance value so that the total resistance is 600 Ω.

Recommended test instruments The internal resistance of the Agilent 3458A is as

follows.

Range

Shunt Resistor

100 mA 1 Ω

▪ Connect the external power supply (24 V ±10) to the Vext terminal as in the figure

below.

The Vext terminals of the left and right connectors are connected inside the module,

therefore the power must only be connected to one of the terminals. Do not connect to
both terminals.

DMM +

-

1

2

3

4

5

6

7

8

9

10

11

12

13

14

15

1

2

3

4

5

6

7

8

9

10

11

12

13

14

15

V+

-

I+

V+

-

I+

V+

-

I+

V+

-

I+

V+

-

I+

V+

-

I+

V+

-

I+

V+

-

I+

CH1

CH2

CH3

CH4

CH5

CH6

CH7

CH8

+

-

+

-

Vext

Vext

DMM +

-

+

-

DC power supply

24V±10%

For voltage measurement

For current mesurement

600Ω ± 5%

2.9 Tests of the 8-CH Analog Output Module

This manual is related to the following products: