Tests, Instruments used – Yokogawa PC-Based MX100 User Manual

Page 83

background image

2-54

SM MX100-01E

2.12 Tests of the 10-CH Screw Terminal Block

(772061)

Tests

2.12.1 Insulation Resistance Test
2.12.2 Withstanding Voltage Test
2.12.3 ID Check
2.12.4 Between-Channel Measurement Error Test
2.12.5 Reference Junction Compensation Accuracy Test

Instruments Used

Instrument

Standards, Recommended Products Etc.

Insulation resistance tester

Kikusui TOS-9201 or equivalent

Withstanding voltage tester

Kikusui TOS-9201 or equivalent

DMM

Agilent Technologies 3458A or equivalent
DCV measurement accuracy: 0.005% of rdg + 1µV
Resistance measurement accuracy: 0.005% of rdg + 1 mΩ

DC voltage generator*

FLUKE Datron 9100 or equivalent
Output voltage 100 VDC or more

Rheostat

ALPHA Electronics ADR-1202 or equivalent
0 Ω ±1 mΩ (short)
Other than 0 Ω, ±(0.006% of setting + 2 mΩ)

Temperature tank standardized at 0°C Coper Electronics ZC-114 or equivalent
Thermocouple

Type T

MX100, MW100, or equivalent

MX100 or MW100 Main Module
MX110-UNV-M10 module
MX150 Base Plate
772062 cable

PC *

Includes software required for testing

Ethernet cable

* Ensure that the DC voltage generator meets the following standards.
However, even if it meets the standards, voltage drops can occur as a result of output

impedance, so the output voltage should always be monitored with a DMM.

Output Voltage

Standard

0 V

±1µV (obtain by shorting)

20 mV

±100 ppm

60 mV

±80 ppm

200 mV

±60 ppm

1 V

±60 ppm

2 V

±60 ppm

6 V

±60 ppm

20 V

±60 ppm

100 V

±60 ppm